SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
When you press
The
deemed stale.

Detection Method

In SA Mode, the X-axis is comprised of data points, also known as "buckets".
The number of data points is specified using the Points setting. Learn more in
"Points " on page
Regardless of how many data points are across the X-axis, each data point
must represent what has occurred over some frequency range and time
interval.
From the frequency span of the measurement, the span of each data point is
calculated as (frequency span / (data points-1)). The detection method allows
you to choose how the measurements in each bucket are displayed.
Figure 9-16 Detection using One Bucket
In the graphic above there is one bucket showing Positive peak, Sample, and
Negative peak Detection methods.
Keysight N9938-90006 User's Guide
different FieldFox receiver frequency bands, so even if the
temperature hasn't changed, an alignment may be required when
one changes the Center frequency.
Hold
The alignment process is stopped, but the last alignment
values are applied to subsequent sweeps.
OFF
The alignment process is NOT performed, and the channel
equalization is deactivated.
ChanEq Align Now
Amptd?
annotation lights up whenever the current channel equalization is
201.
An alignment is performed immediately.
211