SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings

Independent Source/Tracking Generator

For best results, when using the independent source feature, InstAlign and
the battery saver features should be turned off. Refer to
Individual Alignments" on page 208
This feature may require an option. For a comprehensive list, view the FieldFox
Configuration Guide at:
https://www.keysight.com/us/en/assets/7018-06515/configuration-guides/
5992-3701.pdf
A tracking generator, a popular option with Spectrum Analyzers, is a source
which always tracks the SA receiver.
Like a traditional tracking generator, the Independent Source feature can set
the internal FieldFox source to track the SA receiver frequency range. It can
also set the internal source to a CW frequency that is independent of the SA
frequency.
Independent Source can be enabled ONLY when the FieldFox is in SA mode.
To view the internal source, you must connect a cable or device between the
RF Output connector and the RF Input connector.
How to make Independent Source Settings
— Press Measure 1 >
184
Source
Source Enable ON OFF
ON
Turns ON the internal source.
OFF
Turns OFF the internal source.
Source Mode
Choose from the following:
Tracking
The internal source frequency can track the SA receiver.
Center frequency
CW
The source is set to a CW frequency independent from the SA
settings.
Srcw
is annotated on the screen.
Coupled CW
The source is set to a CW frequency, but that frequency
"tracks" the SA receiver Center Frequency settings. Therefore, the
signal always appears at center of the display.
the screen.
Normalize ON OFF
With Source Enable ON and Source Mode = Tracking,
Normalize ON stores the current measurement trace into Trace 4 in View
state as a reference trace. All subsequent trace measurements subtract
this Trace 4 reference trace, and therefore become relative to it.
annotated on the screen. Normalize is automatically turned OFF when the
frequency range or Resolution is changed. Learn more about Trace Display
States in
"Trace Display States (SA Mode)" on page
and
"Preferences" on page
Src
is annotated on the screen
Keysight N9938-90006 User's Guide
"How to Access
770.
Srcp
is annotated on
SrcN
is
201.