Mixed-Mode S-Parameters
In NA Mode ONLY, mixed-mode S-parameters (also known as Balanced
measurements) are available with Option 212.
Because the FieldFox has only two test ports, only reflection measurements are
available. Connect the balanced input or output of your DUT to the FieldFox ports 1
and 2.
For highest accuracy, a Full 2-port calibration is required.
All FieldFox settings and features are supported (except Parameter Conversion) with
mixed-mode S-parameters.
NOTE
Learn more about Balanced Measurements with the FieldFox in the FieldFox
Supplemental Online Help:
http://na.support.keysight.com/fieldfox/help/FieldFox.htm
How to make mixed-mode S-parameter measurements
 Press Measure 1
 Then
 Then choose from the following:
To make balanced reflection measurements at the DUT output, connect the DUT
output to the FieldFox ports.

Parameter Conversion

In NA Mode ONLY, converts the active S-parameter trace to an equivalent
impedance (Z), admittance (Y), or reciprocal 1/S-parameter.
How to select parameter conversions
 Press Measure 1
 Then select an S-parameter
 Then
 Then
 Then choose from the following:
22
Frequency Range for the measurement.
IFBW
Narrower bandwidths require more time to sweep, but lowers trace noise.
measurement. More data points require more time to sweep.
calibration without compromising measurement accuracy.
More
Common reflect/common incident for logical port 1
Scc11
o
Differential reflect/differential incident for logical port 1
Sdd11
o
Differential reflect/common incident for logical port 1.
Sdc11
o
Common reflect/differential incident for logical port 1.
Scd11
o
More
Conversion
and
or
Start
Stop
to select the IF Bandwidth for the measurement.
to select the number of data points for the
Resolution
and
to enter a
Center
Span
FieldFox User's Guide