SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
Channel Measurements
RRC Weighting
RRC, or Root-Raised-Cosign weighting, is offered with Channel Power and
ACPR measurements.
When RRC Weighting is applied to transmitted and received power, the edges
of the channel are 'smoothed' to help prevent interference. To accurately
measure a channel that has RRC weighting, set the same value of RRC
weighting as that used in the transmitter and receiver.
Figure 9-21 Channel power measurement with.22 RRC applied
RRC Weighting is set and enabled automatically when included in a selected
radio standard.
To set and enable RRC Weighting:
1. Press Meas Setup 4 >
2. Enter a value between 0 (no smoothing) and 1 (most smoothing) using the
3. Press Enter.
226
[current setting]
numeric keypad,
filtering is 22.
RRC Weighting ON OFF
arrows, or the rotary knob. A standard level of
Keysight N9938-90006 User's Guide