NA (Network Analyzer) Mode
NA Mode Settings
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
1. Press Sweep 3.
2. Then choose one of the following:

Resolution (Number of Data Points)

Data points are individual measurements that are made and plotted across the
X-axis to form a trace. Select more data points to increase measurement
resolution. However, more data points also takes more time to complete an
entire measurement sweep.
When the Resolution is changed after a calibration is performed, the cal
becomes interpolated. Learn more in
How to set Resolution
1. Press Sweep 3 > Resolution.
2. Then choose from the following:
Using SCPI, Resolution can be set to ANY number of points between 3 and
10001. See the Programming Guide Help:
https://www.keysight.com/us/en/lib/resources/service-manuals/keysight-fi
eldfox-library-help-and-manuals-2153870.html.

Sweep Time

The fastest possible sweep time is always used as the default setting. Use the
Min Swp Time setting to slow the sweep time when measuring long lengths of
cable. Learn more at the FieldFox Supplemental Online Help:
https://www.keysight.com/us/en/lib/resources/service-manuals/keysight-fi
eldfox-library-help-and-manuals-2153870.html.
104
Single
Automatically sets Continuous OFF and causes FieldFox to
make ONE sweep, then hold for the next Single key press. Hold is
annotated in the upper left corner of the display when NOT
sweeping, and changes to an arrow --> while the sweep occurs.
Continuous
Makes continuous sweeps. This is the typical setting
when battery power is not critical.
— You can also use Run / Hold +/- to toggle between Single and
Continuous.
101 |201
|
401
|
601
|
801
"Interpolation *" on page
|
1001
|
1601
|
4001
| 10001.
Keysight N9938-90006 User's Guide
142.