RTSA (Real-time Spectrum Analyzer) Mode—Option 350
RTSA Mode Settings
— Enter 0 to set Auto Trigger OFF or press the Auto Trigger softkey a second
Trigger Hold-Off
Used with External, Video, and RF Burst triggering, sets the hold-off time
between triggers. (Range: 0.000000 to 10.000000 seconds)
Once a trigger occurs, future triggers are ignored until the trigger hold-off time
expires. Setting the trigger hold-off to zero, essentially turns off the trigger
hold-off feature. Also, you can press Trig Hold-Off
Trigger Settings Annotation
Trigger settings are annotated on the FieldFox screen as shown in the following
figure:
— Trig Type (VidT), AutoTrigger (a), Pos Slope (/)
— Horizontal arrow: Trig Level (green horizontal graticule at -35 dBm)

Single / Continuous / Restart / Time Per Div

This setting determines whether the FieldFox measures continuously or only
once each time the
Single to conserve battery power or to allow you to save or analyze a specific
trace.
— To conserve battery power, in addition to using Hold / Single, you must
Keysight N9938-90006 User's Guide
time to turn OFF Auto Trigger.
Single
also, set the System > preferences >Preferences > Battery Saver to On.
For more, refer to the System Settings chapter and
page
770.
or Run / Hold +/- button is pressed. Use Hold /
>
OFF.
"Preferences" on
269