Power Meter (USB/LAN) Mode - Option 302
Troubleshooting
Troubleshooting
This section contains the following:
LAN Sensors
LAN Sensors
If a LAN sensor cable or LAN sensor is not communicating properly with the
FieldFox:
If the FieldFox LAN is removed during the use of the LAN sensor, the
following steps can be taken to re-establish control of the Sensor:
1. Reconnect the LAN sensor to the FieldFox and await confirmation the
2. Once FieldFox has obtained its IP address again, press the Reconnect
If the Sensor LAN is removed during the use of the LAN sensor, the
following steps can be taken to re-establish control over the
measurement:
1. Reconnect the LAN cable to the sensor.
2. Observe the LEDs near the LAN cable connection:
570
"If the FieldFox LAN is removed during the use of the LAN sensor, the
following steps can be taken to re-establish control of the Sensor:" on
page 570
"If the Sensor LAN is removed during the use of the LAN sensor, the
following steps can be taken to re-establish control over the
measurement:" on page 570
FieldFox has obtained its IP address via the connection.
A self-assigned IP address may be shown briefly while FieldFox tries to get
connected to the LAN again after it has been removed. Before proceeding
to the next step, ensure the correct IP has been obtained by the FieldFox.
a. You can view the IP address by selecting:
System >
System Configuration
IP address.
b. Press
Done
to exit the LAN menu.
button:
Press Meas Setup 4 >
in the softkey menu.
The FieldFox will attempt to restore the connection and the settings for the
sensor.
a. The PWR LED will turn on while the Sensor is waking up and
initializing.
>
More
More
LAN Sensor Settings
>
Keysight N9938-90006 User's Guide
>
LAN
to display the current
Reconnect
location
>