Noise Figure Mode (Option 356)
How to Define a Noise Figure Measurement
Keysight N9938-90006 User's Guide
——Recall ENR Table
softkeys and displays a list of stored noise source
files.
——Storage Device [Internal]
softkeys to set the storage location for your files.
Then choose
——Back
to return to previous menu.
T Cold Default
When making measurements at different
ambient temperatures, you can set the T
Kelvin (K) manually (User). When set to Default, the T
value is set to the 296.5 K (Default == Default).
Noise Source Control
control softkey choices. (Active only when ENR Mode ==
Smart (Auto)).
Then
choose:
——Reconnect
FieldFox to a noise source. If one is not found, the
FieldFox displays "Noise Source is NOT
CONNECTED". Allow 10 to 20 seconds for the
instrument to detect the noise source.
——Auto Load ENR OFF ON
enabled the FieldFox automatically searches for a
compatible USB noise source and reloads the
stored calibration and temperature data. When
disabled no existing search is made for USB
noise source data to be loaded.
When Auto Load == ON and compatible USB
noise source is detected (e.g.,
U1832/U1833A/B/C/D) the FieldFox briefly
displays:
"Recalled \temp\U1832A.enr"
When Auto Load == OFF, but a valid USB noise
source is discovered, the following message is
displayed:
"Noise Source Found;ENR file NOT auto
loaded."
——Reload ENR Data
USB noise figure sensor data. This softkey is
inactive/grayed out when a Noise Source is not
detected. When a USB noise source is detected
Opens a menu of file recall
Opens a menu of
Internal
(default),
SD
cold
Opens a menu of ENR noise source
Automatically
reconnects the
When Auto Load is
(Default == ON)
Manually reloads any existing
Card, or USB.
value in
cold
677