SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
For special cases, the individual alignments can be accessible via the front
panel soft-keys using the Preference setting SA - Individual Alignment
Control
when this preference is set to ON: Cal 5
"Preferences" on page
The individual alignments may be:
— Coupled (default) - when they are in the same state, or
— Uncoupled - when at least one of them is in a different state than the other
— Executed on demand, all of them, using the
Each of the individual alignments:
— could be in one of three states:
— can be updated on demand using the corresponding
— The
206
>
On/Off. This enables a soft-key for access to Individual Alignments
770, in
ones
— Off - the alignment is not applied and is not updated automatically
— Auto (default) - the alignment will be updated over time and
temperature as needed, no user interaction required
— Hold - the unit holds the alignment values from the last
alignment-execution and does not execute another alignment
automatically
Refer to
"Individual Alignments" on page
All Alignment
soft-key has the following functionality:
— when the individual alignments are Coupled, the state of all of them
can be changed simultaneously using the
sub-menus
— when the individual alignments are UnCoupled, they will be coupled
when using the
will be simultaneously changed accordingly
>
Individual Align. For more refer to
Chapter 31
System Settings.
Align All Now
207.
All Alignment
soft-key sub-menus and their state
Keysight N9938-90006 User's Guide
softkey.
Align Now
soft-key.
All Alignment
soft-key