IQ Analyzer (IQA) - Option 351
I/Q Analyzer (IQA) Mode Settings

How to modify the I/Q data measurement parameters

The following sections assume you have already selected your I/Q data
measurement type (i.e., Spectrum, Waveform (RF Envelope or I/Q Waveform),
or User Defined).
This section contains the following:

Adjusting the Bandwidth Settings

- Press BW 2.
- Then choose from the following:
There is a significant difference between resolution BW in FFT
measurements and the resolution BW of a traditional "sweeping"
spectrum analyzer (SA) measurement. The resolution BW in FFT
measurements does not directly control the IF BW like a traditional SA
swept measurement. Resolution bandwidth is directly related to the
analysis length (window) and or data acquisition capture length (sweep
time).
298
-
Res BW
(Spectrum only) - Sets the resolution bandwidth of the FFT
analysis. Changing this value changes the FFT window size, analysis
length, and the sweep time (data acquisition capture length). This
softkey is only available when IQA is set to Spectrum (Freq Domain).
When the resolution bandwidth value is set to Auto (default) the
FieldFox automatically sets the optimum resolution bandwidth value.
(Default: 100 Hz, Minimum: 10 Hz, maximum: 100 MHz).
-
Analysis Length
Sets the FFT analysis length that is the length of the
complex I/Q data points (I/Q-pairs) vector that is used for spectrum
analysis, expressed in the number of samples. (Couplings: In Auto,
this setting is directly coupled to the Res BW and to Span. Minimum:
8, Maximum: 524,288. Where each FFT length "point" is equal to 1 I
and 1 Q point. So, 237 is equal to 237 I and 237 Q points).
Keysight N9938-90006 User's Guide