ERTA (Extended Range Transmission Analysis) Mode - Option 209
Configure Measurement Settings
Improve Dynamic Range
The following settings can be made to reduce measurement noise which
improves Dynamic Range.
Increase source power as much as possible while maintaining a leveled source.
How to set Source Power
— Press Meas Setup 4
— Then
— Then
— Then either
— Reduce Attenuation to 5 dB, then turn the Preamp ON. Learn how in
— Decrease the Res BW. Learn how in
— Set Frequency Reference to GPS on both units. This requires Option 307
Source Enable and Source Mode
When Source Enable [OFF] is selected, or when Source Mode [CW or
CoupledCW] is selected, then synchronization ends and the two partner units
revert to general SA (non-tracking) sweeps. However, the units are still in
partnership, and other settings continue to be synchronized such as
Frequencies, ResBw, and Points.
622
Remote Source
Source Power
Max Power
unleveled warning appears. Then decrease power until power is again
leveled.
To further reduce measurement noise, on the Receiver unit:
"Preamplifier Control (Opt 235)" on page
page
185.
(Built-in GPS) and two GPS antennae. The GPS reference can be enabled
for several sweeps. Then if you move to a location without an adequate GPS
signal, Hold-over Mode will occur. Learn more about GPS as a frequency
reference in
"Frequency Reference Source" on page
or
Leveled
- Increase source power until the
160.
"Resolution Bandwidth (Res BW)" on
Keysight N9938-90006 User's Guide