Troubleshooting

Carousel Error

Error Message / Symptom
Carousel Error
44
Possible Cause
Obstructions on or around the
Carousel.
Hopper incorrectly seated on the
Carousel and catching on casework.
Dispense lever interfering with
Hopper Slider.
Hidden obstruction.
Instrument Malfunction
Remedy
Check for, and remove if
necessary, any obstructions
which may be preventing
the Carousel from rotating.
Use the Touch Panel or PC
to continue.
Ensure Hoppers are
correctly seated on the
Carousel (see Chapter 2 -
Loading Hoppers Onto the
Carousel).
Use the Touch Panel or PC
to continue.
Remove the Hopper at the
output position and use the
Touch Panel or PC to
perform a Dispense Cycle
and reset the dispense lever.
Replace the Hopper and
retry.
Remove all Hoppers.
Use the Touch Panel to
rotate the Carousel to check
for correct operation.
Call the Thermo Fisher Scientific
Service Department.