Battery fault
Display in system menu
Charging time too long
The residual charging time was too long
Fault during trickle charging
No battery capacity stored
Stored battery capacity incorrect
Battery voltage = 0.5 V!
The battery is defective
UPS is defective
SM-5811-0000.01 ENG
S5 System • Fault messages and troubleshooting
Description
➜
possible cause/effect
◗
Steps to be taken (please proceed in the order listed)
Battery is a high-drain battery.
The battery charge time was exceeded (internal charge time
limit approx. 3 hours)
Deviation of current measured voltage value from specified
actual value detected internally during trickle charging.
(Internal) fault has occurred during storage of battery capacity.
Battery fault detected.
A software fault has occurred in the UPS.
The following applies to all these messages:
➜
One or both batteries is/are defective
➜
UPS module fault
◗
Replace battery (-ies)
◗
Replace UPS module