Hewlett-Packard Thermal Converters are true rms indicators,
yielding a dc output voltage proportional to the temperature
rise resulting from the input power. The Models
1l050A
and
1l051A
offer an exceptionally flat response and
nearly constant impedance over a wide frequency range. These
characteristics make the thermal converters ideal to check the
response of precision ac voltmeters, oscilloscopes and amplifiers.

Specifications

Maximum input voltage:
1l049A: 3 V rms; 1l050A:
Models l1049A, l1050A, l1051A Thermal Converters
1l049A,
1
V rms; 1l05lA:
0.45
V rms.
Input impedance:
50f ! ±0.15r !
Output voltage for maximum input voltage:
lOr!.
Output impedance:
<

Calibration accuracy

In reference to std.
Frequency range
20
Hz to
20
kHz
within
kHz to
kHz
within
20
50
50
kHz to
1
MHz
within
5
Hz to
20
Hz and
within
1
MHz to
10
MHz
10
MHz to
30
MHz
30
MHz to
60
MHz
60
MHz to
100
MHz
3" wide, H4" high,
Dimensions:
cm).
Weight: net
2.2
oz
(62
g); shipping
Price: HP Model
1l049Ao)(', $125;
HP
11051A'*, $125.
Option
001 *: calibration to
60
Option
002*: calibration to
100
Includes individual calibration report with statement of uncertainty, trace­
*
able to NBS. Options include individual correctional data sheet attached to
calibration report.
to
10
MHz.
7.5
mV de.
Standard
measurement
uncertainty
±0.01%
±0.02%
±0.01%
±0.03%
±0.01%
±0.06%
±0.05%
±0.12%
±0.25%
±0.50%
± 1.50%
(7,6
deep
x
11/2"
4,4
3,8
x
1
lb
(450
g).
HP Model
1l050A*, $125;
MHz, add
$25.
MHz, add
$50.