Executing Measurement
Tips for More Accurate Measurement
Tips for More Accurate Measurement
You can use the following techniques to increase measurement accuracy.
Setting measurement time to 8
Set the measurement time to 8, which provides better measurement accuracy.
For the setup procedure, refer to
Selecting an appropriate measurement range
When making a measurement in the fixed range mode, select the
measurement range so that measurement is performed within the
recommended range (refer to
page 10-172
For the setup procedure, refer to
Using the correction functions
The OPEN correction eliminates the error due to the parallel stray admittance
of the measurement cable and test fixture.
The SHORT correction eliminates the error due to the series residual
impedance of the measurement cable and test fixture.
The LOAD correction eliminates the complex error due to the amplitude/phase
error of the measurement cable and test fixture, scanner, and so on.
For details, refer to
(Executing Correction)" on page
Making stable measurement
In a high-noise measurement environment, you can obtain more reliable
measurement results by performing averaging.
For the setup procedure, refer to
Making measurements using a four-terminal pair
You can use four-terminal pair measurement to eliminate measurement errors.
For details, refer to
page
118
and
Table 10-4 on page
Chapter
"Principle of four-terminal pair measurement" on
248.
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
"Selecting Measurement Time" on page
Table 10-2 on page
10-171,
10-173).
"Selecting Measurement Range" on page
5,
"Preparation for Accurate Measurement
91.
"Setting Up Averaging Count" on page
58.
Table 10-3 on
56.