Instrument State Query; Operation Complete - Agilent Technologies PSA Series User And Programming Manual

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

For example:
Agilent Technologies,E4440A,US00000123,B.02.02
Key Type:
There is no equivalent front-panel key.
Remarks:
An @ in the firmware revision information indicates
that it is proto firmware.
Front Panel
Access:
System, Show System

Instrument State Query

*LRN?
This command is not implemented. Other commands are available for
returning the *LRN data. Use *IDN? to return the instrument model
number, serial number, and firmware version. Use the *SAV/*RCL
commands to save and then return the instrument state information.

Operation Complete

*OPC
*OPC?
The *OPC command sets bit 0 in the standard event status register to
"1" when pending operations have finished. It does not hold off
subsequent operations.
The *OPC? query stops new commands from being processed until the
current processing is complete. Then it returns a "1", and the program
continues. This query can be used to synchronize events of other
instruments on the external bus.
The instrument does not wait for completion of all processes for these
commands. The processes that are monitored are identified in the
STATus:OPERation resgister. These include:
PSA Process
Calibrating
Sweeping
MEASuring (not in all
modes)
Waiting for trigger
Chapter 6
Common Using the STATus System IEEE Commands
STATus:OPER
Byte Value
Register Bit
0
1
3
8
4
16
5
32
Using the STATus System
365