Remote Command:
[:SENSe]:SIDentify[:STATe] OFF|ON|0|1
[:SENSe]:SIDentify:[STATe]?
Example:
SID 1

2.8.8 Signal ID Mode

Displays a menu to select the method of signal identification.
Key Path:
Input/Output
State Saved:
Saved in Instrument State
Factory Preset: Image Suppress
Remote Command:
[:SENSe]:SIDentify:MODE ISUPpress|ISHift
[:SENSe]:SIDentify:MODE?
Example:
SID:MODE ISUP
2.8.8.1 Image Suppress
This signal identification method attempts to suppress all but valid responses by
mathematically removing all image and multiple responses of signals present at the mixer
input. The analyzer internally acquires the data in a two sweep sequence, operates on the
acquired data, and displays the result in Trace 1. Since two measurements are taken for
each display cycle, the display update rate is reduced.
Key Path:
Input/Output
State Saved:
Saved in Instrument State
Remote Command:
See
"Signal ID Mode" on page 161
Example:
SID:MODE ISUP
Chapter 2
,
External Mixer
,
,
External Mixer
Signal ID Mode
Instrument Functions: A - L
Input/Output
161