Instrument Functions: A - L
AMPLITUDE / Y Scale
2.1.11.2 Antenna, Cable, Other, and User Keys
Pressing
,
Antenna
Cable
correction data. These 4 keys display the status of correction sets. If the key indicates
then amplitude corrections for this type have been enabled. To perform the corrections,
both this key and
Apply Corrections
correction set menu located under
Key Path:
AMPLITUDE / Y Scale
Remote Command:
There is no remote command for this key.
2.1.11.2.1 Correction
Turns the amplitude correction function on or off for the specific set of correction data. The
corrections state must be set to
,
NOTE
Antenna
values. This indicates a loss in the external device.
typically entered as negative values which indicate a gain in the external
device.
Key Path:
AMPLITUDE / Y Scale
State Saved:
Saved in Instrument State
Factory Preset: No
Remote Command:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Remote Command Notes: [:SENSe]:CORRection:CSET:ALL[:STATe] must be set to on
for this command to function.
CSET number equivalents to front-panel access definitions are as follows:
CSET or CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
Example:
CORR:CSET2 ON
CORR:CSET2?
50
,
, or
accesses the Correction menu for that type of
Other
User
must indicate
Correction
,
,
More
Corrections
for the correction to be applied.
On
, and
correction factors are generally entered as positive
Cable
Other
,
,
More
Corrections
. The status is toggled in the
Yes
.
User
,
(
Antenna
Cable
,
On
correction factors are
,
, or
)
Other
User
Chapter 2