Agilent Technologies PSA Series User And Programming Manual page 247

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

4.9.2.2.3 Factory
Sets the preset type to "Factory." When you do a preset, all of the factory default
instrument state will be restored. A factory preset switches the analyzer to the Spectrum
Analysis mode and resets the settings of all the modes to the factory defaults (i.e.
Spectrum Analysis Mode with continuous sweep). Refer to
information.
Key Path:
System
Annunciation/
Factory
Annotation:
State Saved:
Survives
Factory
Default:
Mode
Remote Command:
See "Preset Type" on page 245.
SYST:PRES:TYPE FACT defines the type of preset as the factory preset.
Example:
With factory preset selected, use SYST:PRES to do a factory preset.
4.9.2.3 Save User Preset
Saves the current state of the analyzer into the
instrument Preset Type is set to User and you perform a preset operation. After you save a
state here, you must go to the
as the preset state. Refer to
factory-configuration settings.
Key Path:
System
, if the preset type is set to User. See
Preset
Annunciation/
Annotation:
None
Dependencies/
Couplings:
None
SCPI Status Bits/
OPC Dependencies:
None
Remote Command:
:SYSTem:PRESet[:USER]:SAVE
Example:
SYST:PRES:SAVE
Use SYST:PRES:TYPE USER to set factory preset type to "User."
Then use SYST:PRES to do the preset.
Chapter 4
,
,
Power On/Preset
Preset Type
and power cycle, but not saved in Instrument State.
Preset
key and select
Preset Type
"Preset" on page 211
,
Power On/Preset
"Preset" on page 211
state registor for recall when the
User Preset
in order to have this state used
User
key description for the default
"Save User Preset" on page 214
Instrument Functions: P - Z
System
for more
247

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4448aE4446a

Table of Contents