3.5.2 Measure

Switches the analyzer between triggering the current measurement/sweep continuously or
triggering a single measurement. The front panel
single-measurement mode.
Key Path:
Meas Control
State Saved:
Save
Factory Preset: Continuous
Remote Command:
Use :INITiate:CONTinuous OFF|ON See
Remote Command Notes: This command affects sweeping when in the SA mode. It affects
measurments when a measurement has been selected from the MEASure
command subsystem.
When ON, at the completion of each trigger cycle, the trigger system
immediately initiates another trigger cycle.
When OFF, the trigger system remains in an "idle" state until
CONTinuous is set to ON or an :INITiate[:IMMediate] command is
received. On receiving the :INITiate[:IMMediate] command, it will go
through a single trigger cycle, and then return to the "idle" state.
The query INIT:CONT? returns 1 or 0. 1 is returned when the
instrument is continuous triggering. 0 is returned when it is single
triggering.
Example:
INIT:CONT OFF

3.5.3 Pause or Resume

This function pauses the currently running measurement. Pressing
between pausing and resuming your measurement. The key label will toggle between
and
. If an averaged measurement was in progress, the average counter is frozen
Resume
when the measurement is halted
Key Path:
Meas Control
Remote Command:
:INITiate:PAUSe to pause the measurement
:INITiate:RESume to resume the measurement.
Example:
INIT:PAUS
Remote Command Notes: See
page 194
192
"Sweep" on page 232
"Abort the Sweep or Measurement (Remote Command Only)" on
. for more information.
key also puts the analyzer in
Single
.
Pause
will toggle
Pause