CAL? The query performs a full alignment and returns a number indicating
Example:
the success of the alignment. A zero is returned if the alignment is
successful.
4.9.4.3 Frequency Corrections (Remote Command Only)]
Turns the internal frequency corrections on/off.
Not saved in instrument state
Saved State:
Factory Preset: On
Factory
Default:
On
Remote Command:
:CALibration:FREQuency[:STATe] OFF|ON|0|1
:CALibration:FREQuency[:STATe]?
Example:
CAL:FREQ OFF
4.9.4.4 Align Subsys
Accesses the keys to immediately execute an alignment of one of the subsystems (
,
, and
Align IF
Align ADC
started all other operations are stopped and any alignment messages will be visible on the
display.
Key Path:
System
Remote Command:
There is no remote command for this key.
4.9.4.4.1 Align RF
Initiates an alignment on the RF assembly.
Key Path:
System
Annunciation/
Annotation:
Will display a series of pop-up message boxes, indicating an alignment is
in progress.
If a signal is present which interfers with the alignment a message Align
RF 50 MHz signal detected. will be displayed.
SCPI Status Bits/
OPC Dependencies: Status questionable calibration. Bits 3, 11, or 12 will be set if
fails
Chapter 4
). When one of the subsystem alignments is
Align Current Sys Gain
,
Alignments
,
,
Alignments
Align Subsys
Instrument Functions: P - Z
System
,
Align RF
Align RF
253