2.1.11.2.2 Edit
Accesses menu keys that allow you to create and edit an amplitude-correction factor set. It
puts the analyzer into a split-screen mode where the correction data is displayed in a table
under the trace data. Pressing
table from the screen. New points will be applied only after the editor is closed.
Key Path:
AMPLITUDE / Y Scale
Remote Command:
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA
<freq>,<rel_ampl>{,<freq>,<rel_ampl>} Creates an amplitude-correction factor set
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA:MERGe
<freq>,<rel_ampl>{,<freq>,<rel_ampl>} Adds the points with the specified values to
the current amplitude correction data, allowing you to merge correction data. If too much
data is merged, as many points as possible are merged into the existing data and then an
error is reported.
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?
Remote Command Notes: [:SENSe]:CORRection:CSET:ALL[:STATe] must be set to on
for this command to function.
CSET number equivalents to front-panel access definitions are as follows:
CSET or CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
<freq> is the frequency (in Hz) where the correction should be applied;
no unit is allowed in this parameter
<rel_ampl> is the amount of relative amplitude correction (in dB)
needed; no unit is allowed in this parameter
Example:
:CORR:CSET2:DATA 900E6,0.3,1.0E9,0.35,1.3E9,0.2
Chapter 2
while in this menu will exit the menu and remove the
ESC
,
,
More
Corrections
Instrument Functions: A - L
AMPLITUDE / Y Scale
,
(
,
Antenna
Cable
Other
, or
)
User