Installation Note
Agilent PSA Series Spectrum Analyzer
E4446A and E4448A
Option 322 Retrofit Kit
(required to retrofit Option 122,
80 MHz Bandwidth Digitizer)
Option 340 Retrofit Kit
(required to retrofit Option 140,
40 MHz Bandwidth Digitizer)
Part Number E4446-90014
Printed in USA April 2008

Advertisement

loading

Summary of Contents for Agilent Technologies PSA Series

  • Page 1 Installation Note Agilent PSA Series Spectrum Analyzer E4446A and E4448A Option 322 Retrofit Kit (required to retrofit Option 122, 80 MHz Bandwidth Digitizer) Option 340 Retrofit Kit (required to retrofit Option 140, 40 MHz Bandwidth Digitizer) Part Number E4446-90014 Printed in USA April 2008...
  • Page 2 The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Wide Bandwidth Retrofit Kit Wide Bandwidth Retrofit Kit 80 MHz Bandwidth Digitizer Retrofit Kit 40 MHz Bandwidth Digitizer Retrofit Kit Products Affected: PSA E4446A PSA E4448A Serial Numbers: US4611 /US99999999 MY4611 /MY99999999 SG4611 /SG99999999 To Be Performed By: (X) Agilent Service Center ( ) Personnel Qualified by Agilent ( ) Customer Estimated Installation Time:...
  • Page 4 Wide Bandwidth Retrofit Kit 1. The installation of this kit requires that some performance testing be NOTE performed in order to assure the new option is functioning properly. This installation note includes a list of required performance tests. Completing the list of required performance tests does not guarantee the instrument meets all specifications.
  • Page 5 Wide Bandwidth Retrofit Kit Contents Reference Description Agilent Part Designator Number Wide Band Analog IF Assembly E4440-60215 Wide Band Digital IF Assembly E4440-60262 Bandpass Filter, 3.900 GHz 0955-1391 Lowpass Filter, 4.4 GHz 0955-0519 Cable, ribbon, A31 Wide Band Analog IF to E4440-60341 A32 Wide Band Digital IF Cable (65), coax, 100 MHz Ref from A11 Reference Assy J2...
  • Page 6 ❏ Calibration software. Latest software information and downloads available at http://www.agilent.com/find/calibrationsoftware ❏ Test equipment supported by the calibration software. ❏ PSA Series Spectrum Analyzer and Service Guide. This manual is available as E4446AU or E4448AU Option OBW. ❏ Microsoft® Windows® based personnel computer using LAN connection and CD-ROM drive.
  • Page 7: Installation Procedure

    Installation Procedure Installation Procedure Preliminary Hardware Verification With the instrument powered up, press . Look for the System More Show Hardware Option Driver in the list of assemblies. If the Option Driver is listed, proceed with the installation of Option 122 or 140. If the Option Driver is not listed, you must install the Option 123, Switchable Microwave Preselector Bypass retrofit kit before installing this Option 122 or 140 retrofit kit.
  • Page 8 Installation Procedure Figure 1 Instrument Outer Case Removal Installation Note E4446-90014...
  • Page 9 Installation Procedure Remove the Top Brace 1. Refer to Figure 2. Use the T-10 driver to remove the top screws (3) (one screw is under the security label), and the side screws (2) attaching the top brace (1) to the deck.
  • Page 10 Installation Procedure Install the Retrofit Kit Verify Existing Hardware Configuration 1. Locate the A7 Digital IF assembly. It is in the forth card slot from the power supply. Refer to Figure Figure 3 A7 Digital IF Assembly Location 2. If the existing A7 Digital IF assembly has large metal casting covers as shown in Figure 3, this Option 322 or 340 retrofit kit is the wrong kit for your instrument.
  • Page 11 Installation Procedure 5. Refer to Figure 4. Install the A31 Wideband Analog IF assembly into the board slot next to the A7 Digital IF. This slot is the third slot from the power supply. 6. Locate in the kit, and install the A32 Wideband Digital IF assembly next to the A31 Wideband IF assembly just installed.
  • Page 12 Installation Procedure Figure 5 Cable Routing 11.Reinstall the A10 3rd Converter assembly. Re-attach all cables being sure to follow the color code on the board. 12.Locate the cable with the “60” color code in the kit. Connect the SMB end to the A10 3rd Converter J6 (60).
  • Page 13 Installation Procedure 14.Attach the W66 cable to the SMB connector on the A31 Wideband Analog IF assembly. This is the cable you left unattached in step 15.Locate the cable with the “95” color code in the kit. Connect the SMB end of the cable to the A11 Reference assembly's 300 MHZ Out connector.
  • Page 14 Installation Procedure Changes to the RF Section 18.Figure 6 shows the completed installation of replacement filter FL2, the additional filter FL3 and the associated cables. Refer to this figure when completing the rest of the installation, and to determine the location of the components. Figure 6 RF Section Part Locations 19.Locate FL2 in the instrument and completely remove the cable that connects from the...
  • Page 15 Installation Procedure fastener and attach the appropriate hook or loop half to FL2. Press the ribbon cable against the filter. See Figure 28.Locate the two semi-rigid cables and the FL3 filter in the kit. 29.Refer to Figure 6. Install the W78 cable from the top of FL2 to FL3. Connect the W79 cable from the other end of FL3 to A20 Lowband Assembly at J4.
  • Page 16 Installation Procedure Installing the Option Designator and License Keyword NOTE The option designator 122 or 140 and the license keyword must be entered into instrument memory in addition to the correct firmware before the hardware will function. 1. Follow the directions on the Option Upgrade Entitlement Certificate included in the kit.
  • Page 17 Installation Procedure Perform Test Routines using the Agilent Calibration and Adjustment Software The latest software information and downloads are available at http://www.agilent.com/find/calibrationsoftware Calibration Constant Resets Required WB IF Cal Constant Reset Under Test Plan, select Utilities and Calibration Constant Reset. Perform the test listed above.
  • Page 18 15 dB of each other, and at least 20 dB above the noise floor. End of installation. For assistance, contact the nearest Agilent Technologies Sales and Service Office. To find your local Agilent office access the following URL: http://www.agilent.com/find/assist...