4.6 Single

If the analyzer is in continuous sweep mode and not in a measurement (
changes the sweep control to single sweep, and executes a sweep after the trigger condition is
Single
met. If the analyzer is already in single sweep, pressing
trigger condition is met.
Some instrument settings require more than one sweep to complete the measurement (see
,
), or if you have selected a measurement from the functions under the
BW/Avg
Average
key, this function sets the trigger system to be initiated only once. In this case the
MEASURE
trigger condition can be met only once and then all the necessary sweeps will be executed
to make the measurement or complete the averaging function.
With Average on (
BW/Avg
the average again from a count of zero. Sweeps are averaged until N sweeps are then taken
(where N is the average number), and then the sweep is halted.
Single sweep setting is saved in instrument state
State Saved:
Continuous
Factory Preset:
Remote Command:
:INITiate[:IMMediate]
*TRG
Remote Command Notes: Use the :TRIGger[:SEQuence]:SOURce command to select the
trigger source.
For more information,
See also the *TRG command information in
Example:
*TRG
TRIG:IMM
Chapter 4
,
(On)), pressing
Average
see
"Trig" on page 281
Measure
executes a new sweep after the
Single
resets the average trace and starts
Single
.
Chapter 6
.
Instrument Functions: P - Z
Single
,
), pressing
Meas Off
225

Advertisement

Table of Contents
loading