2.1.11 Corrections

Accesses the
Corrections
select which set of correction factors you wish to modify. These frequency/amplitude
corrections will be applied to the displayed data to correct for system losses/gains outside
the analyzer. Four different sets of correction data can be stored.
Key Path:
AMPLITUDE / Y Scale
Remote Command:
There is no equivalent remote command for this key.
2.1.11.1 Apply Corrections
Pressing
Apply Corrections
only be applied for the sets of correction factors whose correction state is set to On. To turn
a set of correction factors on, use the
User menus.
Key Path:
AMPLITUDE / Y Scale
Annunciation/
Annotation:
When
annotation on the left edge of the display, whether or not a correction set
has been turned on using the
Other, or User menus.
State Saved:
Saved in Instrument State
Factory Preset: No
Remote Command:
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Remote Command Notes: To turn On or Off an individual correction set, use:
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]
Example:
CORR:CSET:ALL ON
CORR:CSET:ALL?
Chapter 2
menu keys that allow you to enable the corrections function and to
,
More
(Yes) turns on the amplitude-correction factors. Corrections will
Correction On Off
,
,
More
Corrections
(Yes) is selected, an A will appear in the screen
Apply Corrections
key in the Antenna, Cable, Other, or
(On) key in the Antenna, Cable,
Correction
Instrument Functions: A - L
AMPLITUDE / Y Scale