2.5.5 Limits

Limit lines can be defined to compare the data to your defined limits and indicate a pass or
fail condition.
accesses menus that allow you to create, modify, and change the
Limits
properties of limit lines. There are two limit lines in the instrument.
Key Path:
Display
State Saved:
Not saved in instrument state.
SCPI Status Bits/
OPC Dependencies: No OPC dependencies.
Factory
Default:
Unaffected by system defaults.
History:
Added with firmware revision A.03.00
Remote Command:
:CALCulate:LLINe[1]|2:DATA
<x-axis>, <ampl>, <connected>{,<x-axis>,<ampl>,<connected>}
:CALCulate:LLINe[1]|2:DATA?
Defines the limit line values, and destroys all existing data. Up to 200 points may be
defined for each limit using the following parameters.
can be frequency or time values as specified by the following command:
<x-axis>
:Calculate:LLINe:CONTrol:DOMain.
Frequencies are always in Hz. Time is always in seconds. No unit is
allowed in this parameter.
Range: -30 Gs to +30 Gs for time limits, -3 kHz to +350 GHz for frequency
limits.
amplitude values are always in units of dBm. Up to two amplitude values
<ampl>
can be provided for each x-axis value, by repeating <x-axis> in the data
list. No unit is allowed in this parameter.
Range: -140 dBm to +100 dBm
<connected> connected values are either "0" or "1." A "1" means this point should be
connected to the previously defined point to define the limit line. A "0"
means that it is a point of discontinuity and is not connected to the
preceding point. The connected value is ignored for the first point.