4.2.2 Mode Preset

This key is only available when the preset type is set to User. Press
,
,
On/Preset
Preset Type
A mode preset does not change the mode. It resets only the current mode settings to the
factory defaults.
Key Path:
Preset
History:
Added with firmware revision A.02.00
Remote Command:
See ":SYSTem:PRESet" on page 186.
*RST
The *RST command always does a mode preset, except that it sets the instrument to do a
single sweep/measurement instead of continuous.
Example:
SYST:PRES:TYPE MODE selects the mode type preset
SYST:PRES immediately presets the current analyzer mode

4.2.3 Factory Preset

This key is available only when the preset type is set to User. Press
,
,
On/Preset
Preset Type
A factory preset returns instrument settings to the factory default values. If you are not
already in the spectrum analysis mode, it switches to that mode. A factory preset does not
reset "persistent" functions such as GPIB address, time/date display style, or
auto-alignment state (see
Key Path:
Preset
Remote Command:
See ":SYSTem:PRESet" on page 186.
:SYSTem:PRESet
Remote Command Notes: This command will not reset "persistent" functions such as
GPIB address, time/date display style, or auto-alignment state to their
factory defaults. Use SYSTem:PRESet:PERSistent.
Defaults" on page 231.
Example:
SYST:PRES:TYPE FACT selects the factory type preset
SYST:PRES immediately presets the analyzer to its factory defaults
Chapter 4
.
User
.
Factory
"Restore Sys Defaults" on page
Instrument Functions: P - Z
,
System
Power
,
System
Power
231.)
See "Restore Sys
Preset
213