Instrument Functions: A - L
Auto Couple
2.2.6.1 Auto
Turns the ADC dither to automatic. It then chooses on or off according to which is most
likely to be the best selection, based on the other analyzer settings such as span, resolution
BW and sweep type (FFT or swept).
Key Path:
Auto Couple
State Saved:
Saved in instrument state
Factory Preset: Auto
Remote Command:
Use [:SENSe]:ADC:DITHer[:STATe] AUTO
Example:
ADC:DITH AUTO