Remote Command:
:DISPlay:ANNotation:CLOCk:DATE:FORMat MDY|DMY
:DISPlay:ANNotation:CLOCk:DATE:FORMat?
Example:
DISP:ANN:CLOC:DATE:FORM DMY
4.9.3.3 Set Time
Enables you to set the time of the real-time clock. Enter the time in 24 hour HHMMSS
format.
Key Path:
System
Dependencies/
None
Couplings:
State Saved:
Survives
SCPI Status Bits/
OPC Dependencies: None
Factory
Default:
Not affected by "
Terminators:
Unitless
Default Terminator: None
Hour (HH): 00 to 23.
Range:
Minute (MM): 00 to 59.
Second (SS): 00 to 59.
Remote Command:
:SYSTem:TIME <hour>,<minute>,<second>
:SYSTem:TIME?
SYST:TIME 12,42,00 Sets the clock to 12:42:00 PM.
Example:
4.9.3.4 Adjust Time Setting (Remote Command Only)
Adjust the instruments internal time by the value entered.
Key Path:
System
SCPI Status Bits/
None
OPC Dependencies:
Terminators:
No units are allowed with the command.
DefaultTerminator seconds
Chapter 4
,
Time/Date
and power cycle, but not saved in Instrument State.