Instrument Functions: P - Z
Peak Search
4.1.8.2 Pk Threshold
Specifies the minimum signal level for the analyzer internal peak identification routine to
recognize as a peak. To be considered a peak, a signal must rise above the
value by at least the value specified in
. This applies to all traces and all windows. Press
Peak Excursn
function to hide the threshold line. Applies to
If
is set to Param, it also applies to
Peak Search
Key Path:
Peak Search
Dependencies/
Couplings:
This function is not available when
Amplitude
State Saved:
Saved in instrument state.
Factory Preset: –90 dBm
dBmV, dB µ V, dB µΑ, V, W, A
Terminators:
Default Terminator: dBm
Resolution/
Rounding/
Truncation:
.001 dBm
Knob Increment: amp scale = LOG: 1 % of dB/div or 0.01 (the larger of the two)
amp scale = LIN: 0.1 dBm
Step Key
Increment:
amp scale = LOG: dB/div
amp scale = LIN:
Range:
From the current reference level to the bottom of the display range
Remote Command:
:CALCulate:MARKer[1]|2|3|4:PEAK:THReshold <ampl>
:CALCulate:MARKer[1]|2|3|4:PEAK:THReshold?
Remote Command Notes: CALC:MARK:PEAK:SEAR:MODE must be set to PARameter if
you want peak search to take advantage of the threshold excursion.
Example:
:CALC:MARK:PEAK:THR -60 dB m sets the threshold to -60 dBm.
See the full example for the CALC:MARK:PEAK:SEAR:MODE command
below.
208
Peak Excursn
Next Peak
,
Search Param
.
, then fall back down by at least the
or select another active
ESC
,
, and
Next Peak Left
.
Peak Search
is set to
Y-Axis
Frequency
Peak Threshold
.
Next Peak Right
instead of
Chapter 4