Using the STATus System
STATus Subsystem
Questionable Integrity Positive Transition
:STATus:QUEStionable:INTegrity:PTRansition <number>
:STATus:QUEStionable:INTegrity:PTRansition?
This command determines which bits in the Questionable Integrity
Condition register will set the corresponding bit in the Questionable
Integrity Event register when the condition register bit has a positive
transition (0 to 1). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type:
Factory Preset: 32767 (all 1's)
Range:

Questionable Integrity Signal Register

Questionable Integrity Signal Condition
:STATus:QUEStionable:INTegrity:SIGNal:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Integrity Signal Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.
Key Type:
Questionable Integrity Signal Enable
:STATus:QUEStionable:INTegrity:SIGNal:ENABle <number>
:STATus:QUEStionable:INTegrity:SIGNal:ENABle?
This command determines which bits in the Questionable Integrity
Signal Condition Register will set bits in the Questionable Integrity
Signal Event register, which also sets the Integrity Summary bit (bit 9)
in the Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Key Type:
Factory Preset: 32767 (all 1's)
Range:
380
There is no equivalent front-panel key.
0 to 32767
There is no equivalent front-panel key.
There is no equivalent front-panel key.
0 to 32767