2.4.1.6 Negative Peak
For each interval (bucket) in the trace,
within the interval.
When
is selected, NPk appears on the left side of the display.
Negative Peak
Key Path:
Det/Demod
Remote Command:
See
"Detector" on page
Example:
DET NEG
2.4.1.7 Quasi Peak
Selects quasi peak detection. This is a fast-rise, slow-fall detector used in making CISPR
compliant EMI measurements. Quasi peak detection displays a weighted amplitude using
specific charge, discharge, and meter-movement time constants as described in CISPR
Publication 16.
Key Path:
Det/Demod
Dependencies/
Couplings:
Whenever the quasi peak detector is selected, the resolution bandwidths
available will be the CISPR set described in the tables below. If the FFT
and Sweep setting is Manual:FFT
Table 2-2
CISPR Bandwidths with
Chapter 2
Negative Peak
79.
Band
Start
Frequency
A
10 kHz
B
150 kHz
C & D
30 MHz
detection displays the lowest sample
is grayed out.
Quasi Peak
Set to
RBW
Auto
Stop
Bandwidth
Frequency
150 kHz
200 Hz
30 MHz
9 kHz
1 GHz
120 kHz
Instrument Functions: A - L
Det/Demod
85