Using the STATus System
STATus Subsystem
Questionable Calibration Event Query
:STATus:QUEStionable:CALibration[:EVENt]?
This query returns the decimal value of the sum of the bits in the
Questionable Calibration Event register.
The register requires that the associated PTR or NTR filters be set
NOTE
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
register is cleared.
Key Type:
Questionable Calibration Negative Transition
:STATus:QUEStionable:CALibration:NTRansition <number>
:STATus:QUEStionable:CALibration:NTRansition?
This command determines which bits in the Questionable Calibration
Condition register will set the corresponding bit in the Questionable
Calibration Event register when the condition register bit has a
negative transition (1 to 0). The variable <number> is the sum of the
decimal values of the bits that you want to enable.
Key Type:
Factory Preset: 0
Range:
Questionable Calibration Positive Transition
:STATus:QUEStionable:CALibration:PTRansition <number>
:STATus:QUEStionable:CALibration:PTRansition?
This command determines which bits in the Questionable Calibration
Condition register will set the corresponding bit in the Questionable
Calibration Event register when the condition register bit has a positive
transition (0 to 1). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type:
Factory Preset: 32767 (all 1's)
Range:
376
There is no equivalent front-panel key.
There is no equivalent front-panel key.
0 to 32767
There is no equivalent front-panel key.
0 to 32767
Chapter 6