4.11.3 Line

Sets the trigger to start a new sweep/measurement to be synchronized with the next cycle
of the line voltage.
Key Path:
Trig
Couplings/
Dependencies: Line trigger is not available when operating from a dc power source.
Remote Command:
See
"Trig" on page 281
Example:
TRIG:SOUR LINE selects line triggering.

4.11.4 Ext Front (Ext Trig In)

Sets the trigger to start a new sweep/measurement whenever the external voltage
(connected to EXT TRIGGER INPUT on the front panel) passes through approximately 1.5
volts.
Key Path:
Trig
State Saved:
Saved in instrument state
Factory Preset: 1.5 V
Remote Command:
See
"Trig" on page 281
Example:
TRIG:SOUR EXT to select front panel external triggering.

4.11.5 Ext Rear (Trigger In)

Sets the trigger to start a new sweep/measurement whenever the external voltage
(connected to TRIGGER IN on the rear panel) passes through approximately 1.5 volts.
Key Path:
Trig
State Saved:
Saved in instrument state
Factory Preset: 1.5 V
Remote Command:
See
"Trig" on page 281
Example:
TRIG:SOUR EXT2 to select rear panel external triggering.