2.1.8.9 dB µ V/m
Sets the amplitude units to dB µ V/m. This is a unit specifically applicable to EMI field
strength measurements. In the absence of a correction factor this unit is treated by the
instrument exactly as though it were dB µ V. You must load an appropriate correction factor
using amplitude corrections for this unit to generate meaningful results. Therefore, this
key is grayed out unless one of the corrections are turned on (in
menu) and
Apply Corrections
Key Path:
AMPLITUDE / Y Scale
Dependencies/
Couplings:
If the grayed out key is pressed an error message is generated.
History:
Added with firmware revision A.06.00
Remote Command:
See
"Y Axis Units" on page
Example:
UNIT:POW DBUVM
2.1.8.10 dB µ A/m
Sets the amplitude units to dB µ A/m. This is a unit specifically applicable to EMI field
strength measurements. In the absence of a correction factor this unit is treated by the
instrument exactly as though it were dB µ V. You must load an appropriate correction factor
using amplitude corrections for this unit to generate meaningful results. Therefore, this
key is grayed out unless one of the corrections are turned on (in
menu) and
Apply Corrections
Key Path:
AMPLITUDE / Y Scale
Dependencies/
Couplings:
If the grayed out key is pressed an error message is generated.
History:
Added with firmware revision A.06.00
Remote Command:
See
"Y Axis Units" on page
Example:
UNIT:POW DBUAM
Chapter 2
is set to Yes.
,
,
More
Y Axis Units
41.
is set to Yes.
,
,
More
Y Axis Units
41.
Instrument Functions: A - L
AMPLITUDE / Y Scale
,
Amplitude
Corrections
,
Amplitude
Corrections
45