The standard event status register contains the following bits:
Bit
Description
0
A 1 in this bit position indicates that all pending operations were completed following
execution of the *OPC command.
1
This bit is always set to 0. (The instrument does not request control.)
2
A 1 in this bit position indicates that a query error has occurred. Query errors have SCPI
error numbers from − 499 to − 400.
3
A 1 in this bit position indicates that a device dependent error has occurred. Device
dependent errors have SCPI error numbers from − 399 to − 300 and 1 to 32767.
4
A 1 in this bit position indicates that an execution error has occurred. Execution errors
have SCPI error numbers from − 299 to − 200.
5
A 1 in this bit position indicates that a command error has occurred. Command errors
have SCPI error numbers from − 199 to − 100.
6
A 1 in this bit position indicates that the
the instrument is in local lockout mode.
7
A 1 in this bit position indicates that the instrument has been turned off and then on.
Chapter 6
key has been pressed. This is true even if
LOCAL
Using the STATus System
Using the Status Registers
359