Center Freq - Agilent Technologies PSA Series User And Programming Manual

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Instrument Functions: A - L
FREQUENCY / Channel

2.7.1 Center Freq

Activates the function that sets the center of the displayed frequency range to the specified
frequency.
Key Path:
FREQUENCY
Annunciation/
Annotation:
Center <value> appears in the lower left corner of the display.
Dependencies/
Couplings:
Center frequency and span are coupled to each other.
State Saved:
Saved in instrument state.
Factory Preset:
Model
E4440A
E4443A
E4445A
E4446A
E4448A
Range:
Model
E4440A
E4443A
E4445A
E4446A
E4448A
Remote Command:
[:SENSe]:FREQuency:CENTer <frequency>|UP|DOWN
[:SENSe]:FREQuency:CENTer?
Example:
FREQ:CENT 5 GHZ sets the center frequency to 5 GHz
FREQ:CENT UP changes the center frequency to 5.1 GHz if you use
FREQ:CENT:STEP 100 MHz to set the center frequency step size to 100 MHz
FREQ:CENT?
144
Center Frequency
13.255 GHz
3.355 GHz
6.605 GHz
22.005 GHz
25.005 GHz
Frequency Range
(with Frequency Offset = 0 Hz)
–100.0 MHz
to
–100.0 MHz
to
–100.0 MHz
to
–100.0 MHz
to
–100.0 MHz
to
26.0 GHz
7.2 GHz
13.7 GHz
44.5 GHz
51.0 GHz
Chapter 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4448aE4446a

Table of Contents