Instrument Functions: A - L
AMPLITUDE / Y Scale
2.1.11.2.2.1 Point
Allows you to create or edit an amplitude-correction factor data point. Up to 200 points may be
defined for each set. Enter the point number to be created or edited by using the numeric keypad,
then press
, or use the knob or step keys to move to an existing point. After selecting a point,
Enter
becomes active.
Frequency
Key Path:
AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved:
Not part of instrument state, saved in a corrections file.
Remote Command:
"Edit" on page
51.
See
2.1.11.2.2.2 Frequency
Allows you to enter the frequency value for an amplitude-correction point. Enter the frequency
value by using the numeric keypad. Change the frequency value by using the step keys or the knob.
After entering a frequency,
A frequency coordinate must always be specified for amplitude-correction factors. Amplitude-correction data is
sorted in the table by frequency. The sorting occurs immediately after you have entered the frequency value via
the front-panel.
NOTE
The amplitude correction entered for the lowest frequency will be applied to all frequencies less
than the lowest frequency entered. Similarly, the amplitude correction for the highest
frequency entered will be applied to all frequencies greater than the highest frequency entered.
Key Path:
AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User), Edit
State Saved:
Not part of instrument state, saved in a corrections file.
Remote Command:
"Edit" on page
51.
See
52