2.2 Auto Couple

Coupled functions are functions that are linked/dependent on other functions. Pressing
displays some of the most important coupled analyzer functions.
Auto Couple
An example of a coupled function is Res BW (resolution bandwidth). When Res BW is set to
auto, the Res BW is automatically set based on the span setting.
Coupled functions are affected depending on how they are coupled. For example, Video BW
is coupled to Res BW, so changing Res BW affects Video BW, but changing Video BW does
not affect Res BW. Changing Video BW puts it in manual (instead of auto-coupled). When
Video BW is set to
Man
When a function is in the
one or more functions are manually set so that the amplitude or frequency measurement
becomes uncalibrated, "Meas Uncal" appears on the top right side of the graticule.

2.2.1 Auto All

Auto-couples all coupled functions. If
. Setting any auto coupled function to
Auto
changing the coupling of other functions.
Coupled functions are functions that are linked. When
automatically couples all instrument settings that have auto/man choices. In most cases,
these auto settings give the most accurate measurements and optimum dynamic range.
When a function is coupled, it is in the
state.
Key Path:
Auto Couple
Dependencies/
Couplings:
When
• Resolution BW
• Video BW
• Sweep Time
Frequency
• CF Step
• Attenuation
and
• FFT & Sweep
• PhNoise Opt
& Sweep
• Detector
Max Hold
• Average Type
• ADC Dither
FFTs/Span
Chapter 2
it is unaffected by Res BW changes.
state, a # will appear next to its annotation on the display. If
Man
Auto All
Man
state. When it is uncoupled it is in the
Auto
is selected:
Auto All
couples to:
couples to:
Res BW
couples to:
Res BW
couples to:
in swept spans, to
Span
couples to:
Ref Level
Int Preamp
couples to:
Res BW
(phase noise optimization)couples to:
sweep type
couples to: marker functions;
and
Min Hold
couples to: the marker functions;
couples to:
Sweep Type
is pressed all coupled functions are set to
(manual), uncouples that function without
is pressed, the analyzer
Auto All
and
Span
Span/RBW
and
VBW/RBW
;
;
Video BW
Detector
Res BW
;
;
Ext Amp Gain
Atten Step
and
Span
Avg/VBW Type
Detector
;
;
Span
Res BW
Instrument Functions: A - L
Auto Couple
Man
;
and
Span
Center
in zero span
;
Max Mixer Lvl
;
and
Res BW
Span
FFT
;
;
Average On Off
and
Scale Type
;
and
ADC Ranging
57
;