Instrument Functions: A - L
Auto Couple
• VBW/RBW
• Span/RBW
• Auto Sweep Time
• FFT & Sweep
• ADC Ranging
• Marker Count, Gate Time
NOTE
Marker Trace
Remote Command:
:COUPle ALL|NONE
:COUPle?
Remote Command Notes: sets all the functions to the manual (not coupled) mode. There
is no front-panel key equivalent to the COUPle NONE command.
ALL puts all the functions into the auto coupled mode.
Example:
COUP ALL
2.2.2 FFT & Sweep
Selects the FFT vs. Sweep key functions.
FFT "sweeps" should not be used when making EMI measurements. When an
NOTE
EMI detector is selected,
first, the EMI detectors are grayed out.
Key Path:
Auto Couple
Remote Command:
[:SENSe]:SWEep:TYPE AUTO|FFT|SWEep changes the sweep type to FFT or swept, or it
lets the analyzer automatically choose the type of analysis to use based on current
instrument settings.
[:SENSe]:SWEep:TYPE?
Example:
SWE:TYPE FFT
58
ratio is set to 1.0
ratio is set to Auto
is set to Normal
is set to Auto:Best Dynamic Range
is set to Autorange
is set to Auto
and
have an
Printer
Auto
Manual:FFT
setting, but are not affected by
is grayed out. If
Manual:FFT
.
Auto All
is selected
Chapter 2