Instrument Functions: P - Z

SWEEP

Default Terminator: seconds
in zero span: 1 µ s to 6000s
Range:
in swept spans: 1 ms to 2000s
Remote Command:
[:SENSe]:SWEep:TIME <time>
[:SENSe]:SWEep:TIME?
[:SENSe]:SWEep:TIME:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TIME:AUTO?
Example:
SWE:TIME 500 ms
SWE:TIME:AUTO OFF
4.8.2 Sweep
Switches the analyzer between continuous-sweep and single-sweep mode. Pressing the
front-panel Single key will also put the analyzer in single-sweep mode, then take a sweep.
Key Path:
Sweep
State Saved:
Saved in instrument state
Factory Preset: Continuous
Remote Command:
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
:INITiate:[IMMediate]
Remote Command Notes: This command affects sweep if not in a measurement, and
affects trigger when in a measurement. A "measurement" refers to any of
the functions under the
sweep or single sweep operation when not in a measurement, and
continuous measurement or single measurement operation when in a
measurement.
When NOT in a measurement, this command does the following:
When ON at the completion of each sweep cycle, the sweep system
immediately initiates another sweep cycle.
When OFF, the sweep system remains in an "idle" state until
CONTinuous is set to ON or an :INITiate[:IMMediate] command is
received. On receiving the :INITiate[:IMMediate] command, it will go
through a single sweep cycle, and then return to the "idle" state.