2.2.2.1 Auto: Best Dynamic Range
This function is automatically activated when
tells the analyzer to choose between swept and FFT analysis, with a
Dynamic Range
primary goal of optimizing dynamic range. If the dynamic range is very close between
swept and FFT, then it chooses the faster one.
While
is selected, this key is grayed out. The status of the FFT & Swept selection
Zero Span
is saved when entering zero span and is restored when leaving zero span.
Key Path:
Auto Couple, FFT & Sweep
Saved State:
Saved in instrument state
Remote Command:
[:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge selects the rules to use when
SWE:TYPE AUTO is selected. This setting, combined with your current analyzer setup, is
used to select either FFT or swept mode.
[:SENSe]:SWEep:TYPE:AUTO:RULes?
Example:
SWEep:TYPE AUTO selects the automatic mode.
SWE:TYPE:AUTO:RUL DRAN sets the rules for the auto mode to dynamic
range.
2.2.2.2 Auto: Best Speed
Selecting
Auto: Best Speed
based on the fastest analyzer speed. While
auto-couple settings are kept in memory and are restored whenever leaving
Key Path:
Auto Couple, FFT & Sweep
Saved State:
Saved in instrument state
Remote Command:
[:SENSe]:SWEep:TYPE:AUTO:RULes SPEed|DRANge selects the rules to use when
SWE:TYPE AUTO is selected. This setting, combined with your current analyzer setup, is
used to select either FFT or swept mode.
See "Auto: Best Dynamic Range" on page 59.
Example:
SWEep:TYPE AUTO selects the automatic mode.
SWE:TYPE:AUTO:RUL SPE sets the rules for the auto mode to speed
Chapter 2
Auto All
tells the analyzer to choose between FFT or swept analysis
Zero Span
Instrument Functions: A - L
is selected. Selecting
is selected, this key is grayed out. The
Auto Couple
Auto: Best
.
Zero Span
59