Instrument Functions: P - Z
System
4.9.10.1 Front Panel Test
Used to test the front-panel keys. It shows a list of all the front-panel keys with counter
numbers indicating the number times the key is pressed. Press the
mode and return to the menu.
Key Path:
System
Annunciation/
Annotation:
Text Screen
Dependencies/
Couplings:
The active function is disabled.
Off
Factory Preset:

4.9.11 Restore Sys Defaults

Resets the system settings, including most "persistent" functions, to their factory defaults.
It also does a Factory Preset that resets the analyzer to the Spectrum Analysis Mode. It
does not reset user data such as saved instrument states.
Persistent functions are things such as the GPIB address, time/date display style, and
auto-alignment state. These are parameters that are unaffected by a power cycle or an
instrument preset.
Table 4-2
System Default Settings
Feature
Automatic Alignment On/Alert/Off
Power On Last/Preset
Preset Type Factory/User/Mode
External Amplifier Gain
Input Impedance
GPIB Address
IP address
Host Name
Subnet Mask
SCPI Telnet Port 5023
SCPI Socket Port 5025
SICL Server
Instrument Color Palette
Printer Setup, Language
268
,
Diagnostics
key to exit the test
ESC
Default Setting
Restored?
On
yes
Preset
yes
Mode
yes
0 dBm
yes
50 Ohm
yes
18
yes
10.10.10.10
no
mypsa
no
255.255.0.0
no
On
yes
On
yes
On
yes
Default palette
yes
PCL3
yes
Chapter 4