Agilent Technologies PSA Series User And Programming Manual page 84

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Instrument Functions: A - L
Det/Demod
Remote Command:
See
"Detector" on page
Example:
DET AVER
2.4.1.4 Peak
For each interval (bucket) in the trace, Peak detection displays the highest amplitude
within the interval. Peak detection is used for CW measurements and some pulsed-RF
measurements. For swept analysis, peak detection basically obtains the maximum video
signal between the end of the previous bucket and the end of the current one. Gain is
increased to compensate for the effects of faster sweep rates, to keep the displayed value
accurate. For FFT analysis, the highest spectral amplitude is displayed, even if that peak
amplitude falls between samples of the spectrum computed in the FFT process.
When
is selected, Peak appears on the left side of the display.
Peak
Key Path:
Det/Demod
Remote Command:
See
"Detector" on page
Example:
DET POS
2.4.1.5 Sample
The sample detector displays the instantaneous level of the signal at the center of the
interval (bucket) represented by each trace point.
Sample detection is primarily used to display noise or noise-like signals.
Sample detection is not best for amplitude measurements of CW-like signals for two
reasons. First, the peak response to a signal can occur between samples, so unless the
Span to RBW ratio is lower than usual, the highest sample can be well below the peak
signal amplitude. Second, for the high sweep rates normally used, the peak response of the
RBW filters is up to –0.5 dB. This sweeping error is compensated when using the peak and
normal detectors by changing the overall gain. But the gain is not changed when in the
sample detector, because to do so would cause errors in the response to noise.
When
is selected, Samp appears on the left side of the display.
Sample
Key Path:
Det/Demod
Remote Command:
See
"Detector" on page
Example:
DET SAMP