Instrument Functions: A - L
Display
2.5.5.1.3 Limit Test
Turns the testing of the limit line
limit or margin, PASS LIMIT# or PASS MARGIN# is displayed in green in the upper-left
corner of the measurement area where # is the number of the selected limit line.
Only positive margins are allowed for lower limits and only negative margins are allowed
for upper limits. If the trace is out of the limit or margin boundaries, FAIL LIMIT# or FAIL
MARGIN# is displayed in red. The results for Limit 2 are displayed below those for Limit 1.
Either
or
must be turned on for
Limit
Margin
The color of your screen annotation is dependent on your analyzer settings
NOTE
and may not correspond to the colors described above.
Key Path:
Display
Display
Dependencies/
Couplings:
If either of the limits or margins are turned off, the test cannot be turned
on. That is, if both
automatically.
State Saved:
Not saved in instrument state.
Factory Preset
and *RST:
Off
Factory
Default:
Off (when a limit line is created)
History:
Added with firmware revision A.03.00
Remote Command:
:CALCulate:LLINe[1]|2:STATe OFF|ON|0|1 to turn limit lines on or off.
:CALCulate:LLINe[1]|2:STATe?
:CALCulate:LLINe[1]|2:FAIL?
Example:
:CALC:LLIN:STATE 1 sets limit line 1 test on.
:CALC:LLIN:STATE? responds with the limit line test status.
:CALC:LLIN:FAIL? Queries the status of the limit-line testing. Returns a
"0" if the data passes, and returns a "1" if there is a failure. This query
value is valid only if
100
or
. If the trace is at or within the bounds of the set
On
Off
Test
,
,
Limits
Limit 1
,
,
Limits
Limit 2
and
Limit
or
Margin
to be turned on.
are set to
, then the test is turned off
Margin
Off
is
.
Test
On