Agilent Technologies PSA Series User And Programming Manual page 40

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Instrument Functions: A - L
AMPLITUDE / Y Scale
Remote Command:
[:SENSe]:POWer[:RF]:PADJust <freq>
[:SENSe]:POWer[:RF]:PADJust?
[:SENSe]:POWer[:RF]:PADJust:PRESelector MWAVe|MMWave|EXTernal
[:SENSe]:POWer[:RF]:PADJust:PRESelector?
Example:.
POW:PADJ:PRES MMW
POW:PADJ 100 KHZ
2.1.7 3 - 26 GHz
Selects the preselector for the analyzers microwave frequency bands.
Key Path:
AMPLITUDE / Y Scale
Dependencies/
Couplings:
This function is not available (grayed out) if:
grayed out if microwave preselector is bypassed (see
Microwave Preselector
grayed out if in external mixing. Error 231 is sent.
grayed out if entirely in band 0. Error 208 is sent.
grayed out if entirely in bands 5-6. Error 208 is sent.
History:
Added in revision A.06.00.
Remote Command:
See
"Presel Adjust" on page
2.1.7.1 26 - 50 GHz
Selects the preselector for the analyzers millimeterwave frequency bands.
Key Path:
AMPLITUDE / Y Scale
Dependencies/
Couplings:
This function is not available (grayed out) if:
grayed out if no part of the span is in bands 5 or 6. Error 208 is sent.
grayed out if using external mixing. Error 231 is sent.
History:
Added in revision A.06.00.
40
,
Presel Adjust
.) Error 229 is sent.
38.
,
Presel Adjust
,
Input/Output
Chapter 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4448aE4446a

Table of Contents