Agilent Technologies PSA Series User And Programming Manual page 242

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Instrument Functions: P - Z
System
Remote Command:
:SYSTem:ERRor[:NEXT]?
Remote Command Notes: The :SYSTem:ERRor[:NEXT]? command queries the earliest
entry to the error queue and then deletes that entry.
Example:
SYST:ERR? returns <error number>,<"error string">, for example
-113,"Undefined header".
*CLS clears the entire error queue.
4.9.1.1 Previous Page
Displays the previous page of the Show Errors screen. This key is inactive (grayed out) if
there is no previous page.
4.9.1.2 Next Page
Displays the next page of the Show Errors screen. This key is inactive (grayed out) if there
is no next page.
4.9.1.3 Clear Error Queue
Clears the front-panel error queue from the
Key Path:
System
Remote Command:
*CLS
242
Show Errors
,
Show Errors
display.
Chapter 4

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4448aE4446a

Table of Contents