2.5.5.1.8 Amptd Interp
Allows you to determine how limit trace values are computed between points in a limit
table. The available interpolation modes are linear (
mode is used for both frequency and amplitude, a straight line is used when interpolating
between points in a limit table. This function does not work in zero span (when the
analyzer is in time domain). Refer to the "Note" in
information.
Interpolation modes determine how limit values are computed between
NOTE
points in the limit table. The appearance of a limit trace is also affected by the
amplitude scale, which may be linear or logarithmic.
Key Path:
Display
Display
State Saved:
Persistent; retains settings, even through a power cycle.
Factory Preset
and *RST:
Not affected by preset
Factory
Default:
Log for a new limit.
History:
Added with firmware revision A.03.00
Remote Command:
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE LOGarithmic|LINear
:CALCulate:LLINe[1]|2:AMPLitude:INTerpolate:TYPE?
Example:
:CALC:LLIN:AMPLLINT:TYPE LOG sets limit lines 1 amplitude
interpolation to LOG.
:CALC:LLIN:AMPLLINT:TYPE? responds with the limit line interpolation
type.
Chapter 2
,
,
Limits
Limit 1
,
,
Limits
Limit 2
Instrument Functions: A - L
) and logarithmic (
Lin
Log)
"Freq Interp" on page 105
Display
. If the linear
for more
107