Agilent Technologies PSA Series User And Programming Manual page 81

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Remote Command:
[:SENSe]:DETector[:FUNCtion] NORMal|AVERage|POSitive|SAMPle|NEGative|QPEak
|EAVerage|EPOSitive|MPOSitiv|RMS
[:SENSe]:DETector[:FUNCtion]?
The query returns a name that corresponds to the detector mode as shown by the following
terms:
NORMal
AVERage
POSitive
SAMPle
NEGative
QPEak
EAVerage
EPOSitive
MPOSitive
RMS
Remote Command Notes:
The query returns a name that corresponds to the detector mode. The RMS
selection is an alias which selects the Average detector and Power
Averaging. Therefore, if RMS has been selected, the query will return the
AVER string.
Example:
DET POS
2.4.1.1 Auto
The system selects normal detection as the default, but if a condition arises where a
different type of detection scheme would be better utilized, the system uses the alternate
scheme. For example, when in Auto mode, the Marker Noise function uses Average
detection because the system determines that the average detector has better
speed/variance tradeoffs for noise-type signals.
Refer to
Figure
2-1, which shows a decision tree of how detection type is determined.
Chapter 2
Normal
Average
Peak
Sample
Negative peak
Quasi Peak
EMI Average
EMI Peak
MIL Peak
RMS (alias)
Instrument Functions: A - L
Det/Demod
81