3.3.5 Mkr∆->Span
Sets the start and stop frequencies to the values of the delta markers. The marker is then
set to normal at the center frequency. Only available in
modes, this function is not available if the marker is off, or in
span is
.
Zero Span
Key Path:
Marker ->
Remote Command:
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:SPAN
Remote Command Notes: Select the delta marker mode with
:CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example:
CALC:MARK2:DELT:SPAN sets the start and stop frequencies to the values of
the delta markers of marker 2.
3.3.6 Mkr∆->CF
Sets the delta marker to the center frequency. Only available in
modes, this function is not available if the marker is off, or in
Pair
span is
.
Zero Span
Key Path:
Marker ->
History:
Added with firmware revision A.02.00
Remote Command:
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:CENTer
Remote Command Notes: Select the delta marker mode with
:CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example:
CALC:MARK2:DELT:CENT sets the center frequency to the value of the delta
marker center frequency of marker 2.